DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

T8300-BAC Ver la hoja de datos (PDF) - Amphenol Aerospace

Número de pieza
componentes Descripción
Fabricante
T8300-BAC Datasheet PDF : 8 Pages
1 2 3 4 5 6 7 8
Wall Mount Specifications
Sensing Method
• Non-dispersive infrared (NDIR) absorption
• Gold-plated optics
• Patented ABC Logic self calibration algorithm
CO2 Measurement Range
T8100/T8200/T8300
0 to 2000 ppm (0 ppm = 0 V, 4 mA;
2000 ppm = 10/5V, 20 mA)
T8100/T8200/T8300 - 5P models
0 to 5000 ppm (0 ppm = 0 V, 4 mA;
5000 ppm = 10/5V, 20 mA)
T8200 (10P models) 0 to 10,000 ppm
(0 ppm = 0 V, 4 mA; 10,000 ppm = 10/5V, 20 mA)
T8200 (20P models) 0 to 20,000 ppm
(0 ppm = 0 V, 4 mA; 20,000 ppm = 10/5V, 20 mA)
CO2 Accuracy
± 30ppm or 3% of the reading shown, whichever is higher *
Power Supply Requirements
18-30 VAC RMS, 50/60 Hz, or 10.8 to 42 VDC,
polarity protected
Power Consumption
Typical 0.7 W at nominal voltage of 24V AC RMS
Temperature Dependence
0.2% FS per °C (±0.11% per °F)
Stability
T8100/T8300 - Single Channel
<2% of FS over life of sensor (15 years)
T8200 - Dual Channel
<5% of FS or <10% reading annual over life of sensor
(10 years)
Pressure Dependence
0.135% of reading per mm Hg
Warranty
24 months on mechanical defects
Calibration - lifetime warranty for T8100 and T8300 series
Certifications
CE and RoHS compliant
Signal Update
Every 5 seconds
CO2 Warm-up Time
• < 2 minutes (operational)
• 10 minutes (maximum accuracy)
Operating Conditions
• 32°F to 122°F (0°C to 50°C)
• 0 to 95% RH, non-condensing
Storage Conditions
-40°F to 158°F (-40°C to 70°C)
Flammability Classification
UL94 5VA
Passive Thermistor Type (not Bacnet version)
NTC 10 K W thermistor
Thermistor Accuracy
±1°C (15° to 35°C)
RH Sensing Element
Capacitive polymer sensor
RH Range
0% to 99% RH (non-condensing)
RH Accuracy (25°C)
±2.5% RH (20 to 80% RH)
±3.5% RH (<20% and >80% RH)
Active Temperature Accuracy
±0.8°C @ 22°C
Active Temperature Range
32°F to 122°F (0 to 50°C)
Note: Active analog output standard and only available on
T8100-R and T8100-H versions.
Output
Analog Version
• 0 to 10 V (100 W output impedance) and
• 4 to 20mA (RL maximum 500 W) available simultaneously
BACnet Version
• MS/TP
• RS485
• Baud rates 38400 or 9600
*C02 accuracy statement excludes standard gas used for
calibration that has an accuracy of 2%. In addition, there
is a potential digital to analog error of up to 1%.
3

Share Link: 

datasheetq.com  [ Privacy Policy ]Request Datasheet ] [ Contact Us ]