DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

PBM65 Ver la hoja de datos (PDF) - XP Power Limited

Número de pieza
componentes Descripción
Fabricante
PBM65
XPPOWER
XP Power Limited XPPOWER
PBM65 Datasheet PDF : 2 Pages
1 2
Medical Open Frame
65 Watts
PBM65 Series
THE XPERTS IN POWER
4000 VAC Isolation
Low Leakage Current
Meets Class B, EN55011 Conducted
Medical Safety Agency Approvals
Both Line and Neutral Fused
Fixed Switching Frequency
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Earth Leakage
Current
• 85-264 VAC
• 47-63 Hz
• 1.6 A max at 115 VAC
1.0 A max at 230 VAC
• 17 A max at 115 VAC
40 A max at 230 VAC
• 90 µA max at 115 VAC, 60 Hz
150 µA max at 230 VAC, 50 Hz
Output
Output Voltage
• 5.1 to 28 VDC
Output Voltage
Adjustment
• ±5%
Output Power
• 65 Watts
Minimum Load
• No minimum load required
Line Regulation
• ±0.5% max at full load
Load Regulation
• ±2% typical
Ripple and Noise
• ±1% pk-pk max (See Note 2)
Transient Response • 4% maximum deviation, 500 µs
recovery time to within 1% for a
25% load change
Temperature
Coefficient
• ±0.04%/°C
Hold Up Time
• 10 ms minimum at 110 VAC
Overvoltage
Protection
Overcurrent
Protection
• 112-132% with auto recovery
• 125-200% with auto recovery
Short Circuit
Protection
• Trip & restart with auto recovery
Remote Sense
• Compensates for up to 0.5 V drop
General
Efficiency
MTBF
Switching Frequency •
Isolation Voltage
Size
Weight
Environmental
Operating
Temperature
Cooling
Relative Humidity
Storage
Temperature
EMC & Safety
Safety Approvals
EMI/EMC
ESD
Radiated Immunity •
EFT/Burst
Surge
66% minimum for 5 V models
75% minimum for all others
400,000 hrs min MIL-HDBK-217F
42 kHz ±5 kHz
4000 VAC Input to Output
1500 VAC Input to Output
500 VAC Output to Ground
5.0 x 3.0 x 1.2”
127.00 x 76.20 x 30.48 mm
330 g approx (PCB format)
0 °C to +70 °C
Derate linearly from 100% load at
+50 °C to 50% load at +70 °C
Convection cooled
5-95% non-condensing
-40 °C to +85 °C
UL2601-1, CSA C22.2 No 601.1,
EN60601-1 CE Mark (LVD)
EN55011 (CISPR11) Class B and
FCC Part 18 Level B radiated &
conducted
EN61000-4-2, Level 3 air,
Level 2 contact, Perf Criteria A
EN61000-4-3, 3 V/m, Perf Criteria A
EN61000-4-4, Level 3, Perf Criteria A
EN61000-4-5, Level 3, Perf Criteria A
www.xpplc.com