2 TECHNICAL DATA
CHARACTERISTICS (CONT’D)
Typical response curve
of HS 1100/HS 1101 in humidity
205
200
195
190
Calibration data are traceable to NIST
185
standards through CETIAT laboratory.
180
Measurement frequency : 10kHz
Ta = 25°C
175
170
165
160
0
10
20
30
40
50
60
70
% RH
80
90
100
Polynomial response : C(pf)=C@55%*(1.2510-7RH3-1.3610-5RH2+2.1910-3RH+9.010-1)
Measurement frequency influence
RH in % RH
In this data sheet, all capacitance measurements are @ 10kHz. However, the sensor can operate without restriction from
5kHz to 100kHz. To calculate the influence of frequency on capacitance measurements :
C@fkHz=C@10kHz(1.027-0.01185Ln(fkHz))
Polarization
In order to get a better reproducibility during measurements, always connect the case of the header (pin 2) to the ground
of the circuit.
The case of the header is located on the opposite side of the tab.
Soldering instructions : see the Application Note HPC007
PROPORTIONAL VOLTAGE OUTPUT CIRCUIT
Internal Block Diagram
Reference Oscillator
Sensor Oscillator
L.P. Filter
Gain
Vout
Vout=Vcc*(0.00474*%RH+0.2354)
for 5 - 99% RH
Typical temperature coefficient :
+0.1% RH/°C - From 10 to 60°C
RH
0
Voltage (V)
-
HPC001 Rev. 7 June 2002
DEMO BOARD AVAILABLE ON REQUEST (REF HM1510)
Typical Characteristics for Voltage Output Circuit
At Vcc 5V - 25°C
10
20
30
40
50
60
70
80
90
100
1.41
1.65
1.89
2.12
2.36
2.60
2.83
3.07
3.31
3.55

Share Link: 