Measuring
function
Frequency
Period
Ratio f1/f2
Pulse Width
(Auto Trigger)
Duty Factor
Random Uncertainty rms
Systematic Uncertainty
LSD Displayed
(250ps)2 + (Trigger Error)2
±
x Freq. or Period
Measuring Time
(Prescaler Factor)2+ (f1 x Trigger Error of f2)2
±
f2 x Measuring Time
± (250 ps)2+(Trigger Error)2
± (250 ps)2+(Trigger Error)2 x Frequency
± Time Base Error x Freq. or Period
250ps x Freq. or Period
Measuring Time
± QE x Freq. or Period
Measuring Time
Prescaler Factor
f2 x Measuring Time
± Time Base Error x Pulse Width
± 0.5 x Transition Time ± 1.5 ns
± (0.5 x Transition Time ± 1.5 ns)
x Frequency
100 ps
1 x 10-6
Table 1. Measurement Uncertainties and LSD Displayed
Random Uncertainty
Random uncertainty is due to quantization error, short-term Time
Base stability, internal noise and input signal noise. The random
uncertainty can be reduced by increasing the measurement time.
Trigger Error: Internal noise and input signal noise, expressed as
an rms Trigger Error.
Trigger Error =
1.4 x (eamp)2 + (en)2
Signal slew rate (V/s)
at trigger point
Where:
eamp = rms input amplifier noise (250 µV rms typical)
en = rms noise of the input signal over a 300 MHz bandwidth
Systematic Uncertainty
See crystal oscillator specifications for aging and possible
frequency deviation due to the oscillators temperature
dependency
LSD Displayed
Unit value of Least Significant Digit (LSD) displayed. After
calculation, the LSD value is rounded to the nearest decade
before display (for example >0.5 Hz will be 1 Hz and <0.5 Hz
will be 0.1 Hz). LSD blanking is available to reduce displayed
resolution. Measuring times >1s can give significance in > 10
digits. The 11th and 12th digits can be displayed using the
display overflow function.
Options
Battery Unit (Option PM 9623)
The PM 9623 is a rechargeable battery unit for mounting inside
the counter.
Battery Type:
Sealed lead-acid cells
Battery Capacity:
At 25C
Standby Mode:
Typically 20 hours with
Oven Time Base
Operating Mode:
Typically 3 hours without
options, 2.5 hours with Oven
Time Base, and 2 hours with
Oven Time Base and Input C
Recharge Time:
Battery Protection:
External DC:
Line Failure
Protection:
Temperature
Operating:
Storage:
Weight:
Typically 8 hours in
standby mode
Overcharge and deep
discharge protection
12V to 24V via socket on rear
panel (16V to 24V to charge
internal battery)
Counter automatically switches
to internal battery or external
dc when the line voltage falls
below 90V ac
0°C to +40°C
-40°C to +50°C
1.5 kg (3.3 lb)
GPIB (Option PM 9626/02)
Programmable
All front panel and
Functions:
AUX MENU functions
Compatibility:
IEEE 488.2-1987, SCPI 1991.0
Interface Functions: SH1, AH1, T6, L4, SR1, RL1,
DC1, DT1, E2
Maximum
200 to 1600 readings/s,
Measurement Rate
depending on measurement
to Internal Memory: function and internal data format
Internal
764 to 2600 readings,
Memory Size:
depending on measurement
function and internal data format
Maximum Bus
150 to 1000 readings/s,
Transfer Rate from
depending on internal data
internal memory:
format and output data format
Data Output Format: ASCII, IEEE double
precision floating point
Time Out:
Off or 100 ms to 25.5s in
100 ms steps
Analog Output:
0 to 4.98V in 20 mV steps,
derived from three consecutive
digits selected from the
measurement result
Output Impedance:
200